SPECIFICATIONS | |
Model | ADAMS A1c Lite HA-8380V |
Samples | Whole blood or Hemolysates |
Measured items | HbA1c (stable HbA1c ) và HbF (HbS and HbC can be detected in the Variant mode) |
Measurements range | HbA1c : 3 – 20%, 9 – 195mmol/mol HbF : 0 – 100% |
Measurement principle | HPLC method (Reserved-phase cation exchange chromatography) |
Detection method | Dual-wavelength colorimetry (Wavelengths measured: 420nm / 500nm |
Sampling method | Cap-piercing |
Resolution | 0,1% Ratio, 1mmol/mol |
Throughput | Variant mode: 160 seconds per sample Fast mode: 100 seconds per sample |
Sample volume | Whole blood: Approx. 4μl Anemia: Approx. 8μl Hemolysas: Approx. 350μl |
Min. required sample volume | Blood collection tube: Minimum 10mm away from the bottom Sample cup: 400μl |
Sample container | Blood collection tube: 12,3 or 15mm diameters x 75 to 100mm length Cup : 500μl |
Rack | Arkray racks – 5 samples/rack |
Sample loading capacity | A maximum of 10 samples |
Display | 20 digits x 2 lines LCD |
Built-in printer | Thermal printer, 58mm thermal paper |
Data memory | 300 measurement results (including calibration results) |
External dimensions | 330 (W) x 515 (D) x 485 (H) mm |
Weight | Approx. 35 kg |
Power supply | AC 100 – 240V±10%, 50/60Hz |
Power input | 300VA |
REAGENTS
Reagents | Materials provided | Intended for use | Manufacturer |
Column Unit 80 | 1 unit/box (2500 test/column) | Measure HbA1c in blood using reserve- phase cation exchange chromatography | Arkray Factory Inc.-Japan |
Eluent 80A | 600ml x 4 Packs | Elutes area before A0 and stabilizies the column | Arkray Factory Inc.-Japan |
Eluent 80B | 600ml x 2 Packs | Elutes Variant Hb (HbC and other) | Arkray Factory Inc.-Japan |
Eluent 80CV | 600ml x 2 Packs | Elutes A0 and Variant Hb (HbS) | Arkray Factory Inc.-Japan |
Hemolysis Washing Solution Lite H | 2000ml x 3 bottles | Use this products to dilute whole blood samples and to wash the tubes in which samples flow | Arkray Factory Inc.-Japan |
Calibrator Lite (Freeze dried) | Low: 5ml x 1 bottle High: 5ml x 1 bottle Diluent: 15ml x 1 bottle | Calibrate each time the column is changed | Arkray Factory Inc.-Japan |
extendSURE HbA1c Control | Level 1 : 0,25ml x 4 Level 2 : 0,25ml x 4 | Are disigned as quality control materials to monitor the accuracy and precision of laboratory testing procedures for HbA1c quantitation | Canterbury Scientic New Zealand |
Control Dilution Set 80 | Reconstituent : 15 ml x 1 bottle Diluent 80: 250 mlx 1 bottle | This set dissolves, dilutes commercial available freeze-dried controls and dilutes whole blood sample | Arkray Factory Inc.-Japan |
Washing Solution for tubes | 250ml x 1 bottle | Is used for washing out protein and fat from blood that have adhered to instruments and equipments | Arkray Factory Inc.-Japan |