SPECIFICATIONS | |
Model | ADAMS A1c HA-8190V |
Samples | Whole blood or Hemolysis sample |
Measurement items | HbA1c (stable HbA1c ) và HbF (In Variant mode, HbS, HbC (can be seperated and HbD, HbE can be detecteded) |
Measurement range | HbA1c : 3 – 20%, 9 – 195mmol/mol HbF : 0 – 100% |
Measurement principle | Reserve-phase cation exchange chromatography |
Measurement wavelengths | 420nm / 500nm (dual-wavelength colorimetry) |
Sampling method | Cap-piercing |
Resolution | 0,1% Ratio, 1mmol/mol |
Processing speed | Variant mode: 58 seconds/sample Fast mode: 24 seconds/sample |
Sample consumption | Whole blood: Approx. 8μL Anemia: Approx. 16μL Hemolysates: Approx. 400μL |
Required sample volume | Sample tube: 10mm or above from the bottom of sample tube Sample cup: 400μL or above |
Sample container | Sample tube: (∅ 13 or 15) x 75 -100mm) Sample cup: 500μL or above |
Number of samples loaded | One-way transportation: Maximum of 50 samples Loop transportation: Maximum of 100 samples |
Display | Color LCD (with touch screen) |
Rack | ARKRAY racks |
Built-in printer | Thermal printer with 58mm thermal paper |
Memory capacity | 900 tests results (including calibration results) |
External output | Serial 1 port (Can be optionally used as an Ethernet port) |
Measurement conditions | Temperature: 10 – 30oC Humidity: 20 – 80%RH (No condensation) |
Dimensions | 530 (W) x 530 (D) x 530 (H) mm (Not including the projection portions, eluent packs and hemolysis washing solution bottle |
Weight | Analyzer: Approx. 41kg; Sampler unit: approx. 4kg |
Power supply | AC 100 – 240V±10%, 50/60Hz |
Power consumption | 300VA |
REAGENTS
Reagents | Materials provided | Intended for use | Manufacturer |
Column Unit 90 | 1 unit/box (4000 test/column) | Measure HbA1c in blood using reserve- phase cation exchange chromatography | Arkray Factory Inc.-Japan |
Eluent 90A | 600ml x 4 packs | Elutes area before A0 and stabilizies the column | Arkray Factory Inc.-Japan |
Eluent 90B | 600ml x 2 packs | Elutes A0 and Variant Hb | Arkray Factory Inc.-Japan |
Eluent 90CV | 600ml x 2 packs | Elutes A0 and Variant Hb (HbS) | Arkray Factory Inc.-Japan |
Hemolysis Washing Solution 90H | 2000ml x 3 bottles | Use this products to dilute whole blood samples and to wash the tubes in which samples flow | Arkray Factory Inc.-Japan |
Calibrator 90
| Low : 3ml x 3 bottles High : 3ml x 3 bottles Diluent : 10ml x 3 bottles | Calibrate each time the column is changed | Arkray Factory Inc.-Japan |
extendSURE HbA1c Control | Level 1 : 0,25ml x 4 Level 2 : 0,25ml x 4 | The controls are disigned as quality control materials to monitor the accuracy and precision of laboratory testing procedures for HbA1c quantitation | Canterbury Scientific New Zealand |
Diluent 90 | 250 ml x 1 bottle | This set dissolves, dilutes commercial available freeze-dried controls and dilutes whole blood sample | Arkray Factory Inc.-Japan |
Washing Solution for tubes | 250ml x 1 bottle | Is used for washing out protein and fat from blood that have adhered to instruments and equipments | Arkray Factory Inc.-Japan |